High performance liquid chromatography instrument for analytical HPLC

The instrument is designed for the reliability, safety and ease of use required for routine pharmaceutical tasks, quality control, and environmental and food analysis. HPLC system consists of Vialsampler G7129A with a high capacity of 132 positions for 2 ml vials. Quaternary pump VL G7111A has an extended voltage range, reaching pressures up to 400 bar and a wide flow range (up to 10 mL/min). This pump has ability to use up to four solvents for the analysis. Detector DAD WR G7115A with a data acquisition rate of 120 Hz and analysis at wavelengths of 190-950 nm, provides detection with low detection limits. Simultaneous detection is possible for up to eight compound-specific wavelengths. The system also consists of a thermostat for one column with choices for 4.6 mm and 3.0 mm ID columns. Data analysis is processed using an OpenLAB CDS program.

What information you can obtain?

  • Qualitative analysis
  • Quantitative analysis (amount of analyte in the sample)
  • Verification analysis

What samples can be measured?

  • substances absorbing in UV-VIS range and soluble in water or liquid solvents

Possible reports:

  • Retention time
  • Peak height
  • Peak area
  • Amount/Concentration
  • LOD and LOQ of used method

Columns

  • Kinetex 5 mm C18 100 x 6 mm (Phenomenex)
  • Kinetex 2.6 mm Polar C18 100  x 30 mm (Phenomenex)
  • Poroshell 4 mm 120 EC‑C18 150 x 4.6 mm (Agilent)

Prep columns

  • SecurityGuard ULTRA Holder for 2.1 to 4.6 mm UHPLC columns (Phenomenex)
  • SecurityGuard ULTRA Cartridges for C18 4.6 mm UHPLC columns (Phenomenex)